The quality problem could lead to release of particles that could block blood vessels.
The action affects vascular guidewires made of nickel-titanium alloy with a radiopaque metallic coil tip. Physicians use the guidewires in minimally invasive procedures to access peripheral blood vessels. The guidewires have a hydrophilic coating.
In some cases, the hydrophilic coating can swell inconsistently when hydrated with saline. Inconsistent swelling, which can lead to clumping of the coating, happens most frequently on the distal coil section of the guidewire. Clumping raises the risk that the coating will flake, become displaced or detach.
If the coating separates from the guidewire, particles could enter the bloodstream and block vessels. The potential for blockages depends on the size and quantity of the released particles. Hydrophilic polymer embolization is a known risk for catheters and other coated intravascular devices. Blockages in the heart, lungs, brain, kidneys, grafts and compromised peripheral circulation pose particular threats.
Potential outcomes include stroke, organ failure and death. When guidewires are used to access veins, coating particles are most likely to reach vessels supplying the lungs. Particles released in a vein could reach the arteries of patients with a right-to-left shunt, an opening that lets blood flow directly between the two sides of the heart.
Heraeus, which does business as Heraeus Medevio, wrote to customers about the issue on Nov. 4 and Nov. 5. The company asked customers to review their inventory and distribution records, immediately stop distributing and quarantine affected lots, and notify any downstream customers who received the devices.
Source link







